| NTRODUCTION This instrument is a digital thermometer for use with any K-type thermocouple as temperature sensor, Temperature indication follows National Bureau of Standards and IEC584 temperature/voltage tables for K-type thermocouples. SPECIFICATIONS ELECTRICAL Measurement Range: -50~+1300℃(-5~+1999°F) Resolution: 0.1℃, 1℃, 0.1°F, 1°F Maximum voltage at thermocouple input:60V DC,24V AC RF Field Derating:Strong RF fields can adversely affect measurement accuracy. ENVIRONMENTAL Ambient operating Range : 0~50℃ (32~122°F) Storage temperature: -40~+60℃ (-40~+140°F) Humidity: 0%~90%(0~35℃) 0%~70%(35~50℃) Basic Accuracy: (@23 ±5℃ Calibration) Function Range Resolution Accuracy ±(% of reading +degrees) ℃ -50.0~199.9℃ 0.1℃ 0.2% ± 1℃ °F -50.0~199.9°F 0.1°F 0.2% ± 1°F ℃ -50.0~1000℃ 1℃ 0.3% ± 1℃ °F -50.0~1999°F 1°F 0.3% ± 2°F ℃ 1000~1300℃ 1℃ 0.5% ± 1℃ NOTE The basic accuracy specification does not include the error of the probe. Please refer to the probe accuracy specification for additional details. Temperature Coefficient: 0℃ to 18℃(32°F to 64.4°F) and 28℃ to 50℃(82.4°F to 122°F) ambient multiply the basic accuracy specification by 0.1 for each degree above 28℃(82.4°F) or below 18℃(64.4°F) GENERAL Weight :157 gm (5.54oz) Dimension:5.31”Lx2.83”W x 1.22”vH(135x72x31mm) Battery:Standard 9V battery (NEDA 1604 ,6F22 or 006P) Battery life:200 hours Low Battery indication:the (BT)is displayed when the battery voltage drope below the operating voltage. Over Range indication:Highest digit of (1) or (-1) is display Measurement Rate:2.5 measurements per second. Display: 3 1/2digit liquid crystal display(LCD) with maximum reading of 1999. Accessories: battery. Instruction Manual. |